MED Probe (MEA)
Microelectrode array (MEA) with the industry’s lowest-impedance electrodes
The MED Probe has multiple microelectrodes embedded in the center of a transparent glass plate. The surrounding glass (or plastic) cylinder makes the MED Probe a self-contained recording chamber.
The MED Probe features the industry’s lowest-impedance electrodes (10 kΩ at 1 kHz for 50 μm electrodes). It provides excellent signal-to-noise ratio and capabilities of stimulation.
Specifications | |
---|---|
Substrate/Chamber | |
Substrate | Glass, T = 0.7 mm |
Conducting layer | Indium tin oxide (ITO), 0.15 μm |
Insulation layer | Acrylic resin (to be replaced with Polyimide when inventry runs out) |
Cylindrical chamber | Glass (OD:25 / ID:22 mm) Acrylic resin (for Multi-well Probe) |
Substrate size | |
MED Probes (all types) | 50 x 50 mm |
MED Probe 16 | 30 x 40 mm |
MED Multi-well Probe | 100 x 35 mm |
Recording Electrodes | |
Material | Platinum Black |
Impedance | |
50 μm electrodes | 10 kΩ (Typ.) |
20 μm electrodes | 15 kΩ (Typ.) |
(1 kHz, 50 mV applied sinusoidal wave) | |
Maximum Stimulus Output | |
Current-driven | ±200 μA (t=300 μS) |
Voltage-driven | ±1V |
Reference Electrodes | |||
Material | Platinum Black | ||
Type | Size | Number of electrodes / chamber (well) | Total impedance (Typ) / chamber (well) (f=1 kHz) |
A | 100 μm | 4 | 550 Ω |
B | 200 μm | 4 | 550 Ω |
B (1) | 200 μm | 1 | 2.2k Ω |
B (2) | 200 μm | 2 | 1.1k Ω |
C | φ100 μm | 16 | 550 Ω |
D | φ100 μm | 16 | 550 Ω |
D (2) | φ100 μm | 8 | 1.1k Ω |
E | 200 μm | 16 | 300 Ω |